Ultrasonic Air Detector Medical Sensor Medical Equipment Infusion Pump Signal Sensor 5VDC 12VDC 24VDC Measurement 6-8MM Tube

Suitable for detecting air, air bubbles, microbubbles or foam in flowing liquids in pipelines

Classification:

Product

medical sensor


  • Product Detail
    • Commodity name: Ultrasonic Air Detector Medical Sensor Medical Equipment Infusion Pump Signal Sensor 5VDC 12VDC 24VDC Measurement 6-8MM Tube
    • Commodity ID: ZBU06

    working principle:
    According to the different propagation characteristics of ultrasonic waves in different media, through special circuit analysis and processing, to
    Intelligently identify the medium condition in the monitored flow pipe

    Features:

    1. High precision, strong anti-interference ability, can detect at least 150ul single small bubble, response time 500μS, detection threshold and response time can also be set according to user requirements;

    2. The snap-in tube installation does not need to cut off the pipeline, and has no pollution to the detection liquid;

    3. Not affected by changes in fluid color and pipe material, it can detect air bubbles in most liquids;

    4. Low cost, long life, low power consumption, maximum 45mA, 5VDC power supply;

    5. This model is suitable for the infusion tube with an outer diameter of 6-8mm (hose), other diameters or rigid pipelines can be customized according to user requirements, and it is suitable for a wide range of applications;

    6. TTL signal output, maximum load 10mA, more control output can be customized according to user requirements;

    7. Provide an "enable" control terminal to freely control the working time of the sensor (optional).

    Specifications: (default style)

    Power supply: 5VDC;

    Output signal (TTL level signal): low level (0V) - gas,


    High level (5V) - liquid.

    Wiring Instructions:

    Red - the power supply is positive;

    Blue - power ground;

    Yellow - Signal.

    Optional customization:

    Power supply: 5VDC /12VDC /24VDC

    Signal output: TTL / NPN / PNP / optocoupler output

    Maximum load: TTL - within 10ma

    NPN—within 50ma